d. ACA
Range Resolution Accuracy Overload Protection
400µA 0.1µA ±0. 8% of reading, Fuse
4000µA 1µA ±3 digits 0.5Amp/600V
40mA 0.01mA
400mA 0.1mA
e. OHM (Resistance,
)
Range Resolution Accuracy Overload Protection
400 0.1 ±0. 5% of reading, 600V DC or
4k 0.001k ±3 digits AC Peak
40k 0.01k
400k 0.1k
4M 0.001M ±1% of reading,
40M 0.01M ±3 digits
f. Diode Test
Test Voltage Max Test Current Over Load Protection
3V Approx. 30µA 600 V DC or Peak AC
g. Continuity Buzzer
Test Voltage Threshold Over Load Protection
3V < 50 600 V DC or Peak AC
h. General Specifications
Max. Volt. between 1000V
any Input and Ground
Fuse Protection mA: 0.5Amp/600VAC
Display Type 4,000 Count, 2 times per second update
Operating Temp. 0° to 40°C (32° to 104°F)
Storage Temp. -10° to 50°C (14° to 122°F)
Relative Humidity 0% to 80%
Power Supply 2 Each 1.5 Volt “AA” Batteries
Battery Life 200 hrs. Typical
Size (H x L x W) 33mm x 86mm x 187mm (1.3in x 3.4in x 7.4in)
Weight 340g (12oz)
3. Specifications
IEC 1010 Over Voltage:
CAT II - 1000V
CAT III - 600V
Pollution Degree 2
a. DCV
Range Resolution Accuracy Impedance
400mV 0.1mV ±0.3% of reading, 10M
4V 0.001V ±3 digits
40V 0.01V
400V 0.1V
1000V 1V
b. ACV (45 to 450 Hz)
Range Resolution Accuracy Impedance
4V 0.001V ±0.8% of reading, 10M
40V 0.01V ±3 digits
400V 0.1V
750V 1V
c. DCA
Range Resolution Accuracy Overload Protection
400µA 0.1µA ±0.8% of reading, Fuse
4mA 0.001mA ±3 digits 0.5Amp/600V
40mA 0.01mA
400mA 0.1mA
9
UL 3111-1
*Warning: Use only correct size, voltage and current rated fuses.
Test Leads: Use only correct type and overvoltage category rating.